Japanese Site

Thai Site

CONTACT

HD500 series | Ultrasonic Level meter|HD500-C / HD500-D
[Measurement Equipment [ Level meters ]]

Overview

  • Hiring high frequency ultrasonic system enables small dead zone and is possible to measure short distance.
  • High resolution, 0.1 mm
  • Minimize the main unit and sensor unit, so easy to install to the limited area.
  • Sensor part is chemical proof plastic (PFA).
  • Equipped RS-232C interface, 4-20 mA current output, Alarm output allows to communicate with existing system.
  • Flow measurement by measuring catchment level is possible.

Usage
Sensing or positioning goods on production line
Liquid level sensing in small size tank



Specification

Model No. HD500-C HD500-D
Number of channels 1
Frequency
200 kHz 400 kHz
Measurement target Liquid
Measurement distance range 0.12 to 1.0 m 0.06 to 0.45 m
Resolution Measurement 0.1 mm
Display
0.1 mm
Accuracy
±0.25 % F.S.
(±0.25cm)
±0.25 % F.S.
(±0.1cm)
Data update cycle 0.05 seconds
Sensor directivity angle 10° (-6dB)
7° (-3dB)
5° (-6dB)
4° (-3dB)
Power source Voltage 12 to 24 V DC ±15%
Power consumption 3 W (500mA when started)
Display 4-digit LED
Output Alarm 2 points each for upper/lower
30V DC, 0.1 A (NPN open collector)
4-20mA current output Resolution:16bits
(Max. load resistance 450Ω)
Interface RS-232C(Transmission distance : Max.10m)
ModelMain unit Sensor unit
Ambient operating temperature
0 to 50℃
Material ABS PFA
Cable: FEP
Capcon : PVDF
Capcon inner seal: PPE-V
Structure IP43 equivalent IP65 equivalent
Dimensions
(WxDxHmm)
113x52.5x94 dia 42x39
Sensor cable length - 2 m
Max. sensor cable length - 2 m
Weight
300 g
150 g
Screw
-
M32 P1.0
●Only HD500-C is standard-equipped with the weir type flowmeter function.
The flow rate can also be measured for triangular weir.




Industrial Equipment TEL:81-532-41-2774